Health Products

IGG-2(TINA-QUANT IGG GEN.2) FOR USE ON C701 ANALYZER (2017-10-10)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Roche Diagnostics GMBH
Sandhoferstrasse 116
Mannheim
68305
GERMANY
IGG-2(TINA-QUANT IGG GEN.2) FOR USE ON C701 ANALYZER5220718190

Introduction

Affected products

IGG-2(TINA-QUANT IGG GEN.2) FOR USE ON C701 ANALYZER

Reason

Following internal investigation, the manufacturer identified a situation where the specifications for QC recoveries could not be fulfilled for the sensitive applications (urine/CSF) for the Tina-Quant IgG Gen-2 assay (cat #05220718190 lot 26358701) on cobas c701-702 analyzers. This situation has no impact in Canada since the identified lot will not be distributed to users.