ALINITY I PROCESSING MODULE (2017-10-13)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: Healthcare Professionals, General Public, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
AI01072ALINITY I PROCESSING MODULE
Manufacturer
Abbott GmbH and Co. K.G.
Max Planck Ring 2
Wiesbaden
65205
GERMANY
03R65-01

Introduction

Affected products

ALINITY I PROCESSING MODULE

Reason

Abbott has identified the potential for a loose cable connection on the reagent cooler, which could result in temperature errors and eventual reagent cooler failure. The loose cable connection has the potential to heat up over long periods of time and cause damage to the wire and connector. This would be detected by the operator in the form of a burning smell. The cable connection to the reagent cooler is located within the processing module and is not accessed by the operator. Due to the very low flammability rating of the materials used, the heat damage is localized to the connector.