Health Products

BC-THROMBIN REAGENT (2017-10-12)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: Healthcare Professionals, Hospitals, General Public
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
46751
Manufacturer
Siemens Healthcare Diagnostics Products Gmbh
Emil-Von-Behring-Str. 76
Marburg
35041
GERMANY
BC-THROMBIN REAGENTOWNA11

Introduction

Affected products

BC-THROMBIN REAGENT

Reason

The BC Thrombin Reagent Kit lots 46751 (contains Thrombin Reagent lot 517468) and Kit lot 47184 (contains Thrombin Reagent lot 517469) produce unexpected prolonged thrombin time results for expected normal samples, and may recover above the upper limit of normal which is 21 sec as stated within the IFU (instruction for use) and the BCS/ BCS XP System application sheets.