Health Products

Cook Percutaneous Enteral Feeding Catheters (2017-10-19)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: Hospitals, General Public, Healthcare Professionals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
G08417
Manufacturer
Cook, Inc.
750 Daniels Way
Bloomington
47404
Indiana
UNITED STATES
B. WILLS-OGLESBY PERCUTANEOUS GASTROSTOMY SET MALLINCKRODTWOGS-1400-MKDT-A
G05274
Manufacturer
Cook, Inc.
750 Daniels Way
Bloomington
47404
Indiana
UNITED STATES
A. WILLS-OGLESBY PERCUTANEOUS GASTROSTOMY SETWOGS-1200-WH-050493

Introduction

Affected products
  1. WILLS-OGLESBY PERCUTANEOUS GASTROSTOMY SET
  2. WILLS-OGLESBY PERCUTANEOUS GASTROSTOMY SET MALLINCKRODT

Reason

Cook has identified that the affected lots may contain wire guides that were incorrectly loaded into the wire guide holder. This could lead to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip.