Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 100 numbers, contact manufacturer
Manufacturer
Draeger Medical Infant Care, Inc.
3135 QUARRY RD,
TELFORD
18969
Pennsylvania
UNITED STATES
DUAL HEMO MCABLE PODSMS20783

Introduction

Affected products

DUAL HEMO MCABLE PODS

Reason

Draeger Medical Systems inc.  has determined through internal testing that Dual Hemo MCable Pods with Revision Index (RI) 15 and 16, which are used with the Draeger Infinity Acute Care System, may permit liquid ingress inconsistent with their IPX4 rating, which could potentially result in an incorrect measured value of invasive blood pressure.