Monteris Medical NeuroBlate system (2017-10-05)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: Healthcare Professionals, General Public, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
All lotsMonteris Medical NeuroBlate system
Manufacturer
MONTERIS MEDICAL CORPORATION
14755 27TH AVE NORTH, SUITE C
PLYMOUTH
55447
Minnesota
UNITED STATES
NB102

Introduction

Affected products

Monteris Medical NeuroBlate system

Reason

After reports of probe damage during four procedures, the investigation concluded a 1.5 MRI system interacted with the NeuroBlate probe such that radiofrequency (rf) energy was transferred to the probe, resulting in unintended heating and damage at the tip of the probe.