A. EtCO2 OxyMask
B. OxyArm- Oxykid
C. OxyArm-OxyPlus
Reason
On October 6, Southmedic received a customer complaint in which the CO2 diffuser of the EtCO2 mask became disconnected from the mask and fell into the patient's mouth. Further inspection of the identified lot number and investigation into root cause identified the potential for loose or missing diffusers in other lots.