IPRO CARELINK THERAPY MANAGEMENT SOFTWARE (2017-10-10)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: Healthcare Professionals, General Public, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
3030864IPRO CARELINK THERAPY MANAGEMENT SOFTWARE
Manufacturer
MEDTRONIC MINIMED
18000 DEVONSHIRE ST
NORTHRIDGE
91325
California
UNITED STATES
MMT-7340

Introduction

Affected products

IPRO CARELINK THERAPY MANAGEMENT SOFTWARE

Reason

Due to a time conversion error, data uploaded from the iPro2 recorder to the CareLink iPro software using the new CareLink iPro uploader feature may result in incorrect reports. This pertains only to reports generated between September 23, 2017 and September 27, 2017 that used the new CareLink iPro uploader feature. Specifically, when data from the iPro2 recorder was uploaded using the new CareLink iPro uploader feature, the glucose reading timestamps inadvertently defaulted to Greenwich Mean Time (GMT) /Coordinated Universal Time (UTC). Manufacturer determined that the difference in time zones could result in calculation and display errors on the reports. This could lead to potential differences or inaccuracies in the sensor tracings, which also affect the display of meals and activities and the resulting report.