Health Products

PLUM 360 INFUSION SYSTEM WITH MEDNET (2017-10-19)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: Healthcare Professionals, Hospitals, General Public
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
SOFTWARE V15.10.00.010PLUM 360 INFUSION SYSTEM WITH MEDNET
Manufacturer
ICU Medical Inc.
600 NORTH FIELD DRIVE
LAKE FOREST
60045
Illinois
UNITED STATES
30010-XX-YY

Introduction

Affected products

PLUM 360 INFUSION SYSTEM WITH MEDNET

Reason

Issue 1: The user is not able to turn off the pump when a pump in a distal occlusion paused state is exposed to a second "power off then on" malfunction.
Issue 2: When the user accesses the preventive maintenance screen, in service/biomed mode, on Plum 360 v15.10.00.010 devices, with a total delivery time greater than 1,500 hours, the user will not be able to interact with the device and the device must be power cycled.