B. STERILMED REPROCESSED STRYKER EXTERNAL FIXATION SYSTEMS
STR4920-1-010
All lots
Manufacturer
STERILMED, INC.
5010 CHESHIRE PARKWAY, SUITE 2
PLYMOUTH
55446
Minnesota
UNITED STATES
C. STERILMED REPROCESSED SMITH & NEPHEW EXTERNAL FIXATION SYSTEMS
SMI10-0105
All lots
Manufacturer
STERILMED, INC.
5010 CHESHIRE PARKWAY, SUITE 2
PLYMOUTH
55446
Minnesota
UNITED STATES
A. STERILMED REPROCESSED DEPUY SYNTHES EXTERNAL FIXATION SYSTEMS
SYN03.311.011
Introduction
Affected products
A. STERILMED REPROCESSED DEPUY SYNTHES EXTERNAL FIXATION SYSTEMS
B. STERILMED REPROCESSED STRYKER EXTERNAL FIXATION SYSTEMS
C. STERILMED REPROCESSED SMITH & NEPHEW EXTERNAL FIXATION SYSTEMS
Reason
Sterilmed has initiated a medical device correction for all lots of listed Sterilmed reprocessed DePuy Synthes, Stryker, and Smith & Nephew external fixation systems. Sterilmed has discovered the product label and IFU do not contain information regarding the safety and compatibility of the subject products in a magnetic resonance environment and such labeling is inconsistent with the OEM labeling.