Health Products

SICD SYSTEM - Q-TECH PROGRAMMER (2017-07-04)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
All lotsSICD SYSTEM - Q-TECH PROGRAMMER
Manufacturer
Cardiac Pacemakers Incorporated
4100 Hamline Avenue North
St. Paul
55112
UNITED STATES
3200

Introduction

Affected products

SICD SYSTEM - Q-TECH PROGRAMMER

Reason

Observations indicate that this device delivered an atypical amount of energy likely contributing to patient death. The most likely cause was a transient corruption of the S-ICD operating state or device memory due to a Single Event Upset (SEU). SEU's are known to affect memory in electronic devices and high reliability devices (pacemakers or defibrillators) include mechanisms to deal with this issue to reduce the occurrence of potentially harmful malfunctions.