Introduction
Affected products
A. ANACONDA LONGER LEG ILIAC STENT GRAFT SYSTEM (STRAIGHT)
B. ANACONDA LONGER LEG ILIAC STENT GRAFT SYSTEM (FLARED)
C. ANACONDA LONGER LEG ILIAC STENT GRAFT SYSTEM (TAPERED)
Reason
Vascutek has received two complaints whereby a small section of the delivery system sheath became detached from the collar during deployment of the Anaconda Longer Leg Iliac Stent device from the delivery system. Should this occur users should refer to the bailout procedure in the instructions for use (IFU) to complete the deployment.