Health Products

ANACONDA LONGER LEG ILIAC STENT GRAFT SYSTEMS (2017-11-09)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: Healthcare Professionals, General Public, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
VASCUTEK LIMITED A TERUMO COMPANY
NEWMAINS AVENUE
INCHINNAN, GREAT BRITAIN
PA4 9RR
B. ANACONDA LONGER LEG ILIAC STENT GRAFT SYSTEM (FLARED)FL1213X150
2002224604
Manufacturer
VASCUTEK LIMITED A TERUMO COMPANY
NEWMAINS AVENUE
INCHINNAN, GREAT BRITAIN
PA4 9RR
C. ANACONDA LONGER LEG ILIAC STENT GRAFT SYSTEM (TAPERED)TL1210X150
2002073780
Manufacturer
VASCUTEK LIMITED A TERUMO COMPANY
NEWMAINS AVENUE
INCHINNAN, GREAT BRITAIN
PA4 9RR
A. ANACONDA LONGER LEG ILIAC STENT GRAFT SYSTEM (STRAIGHT)L12X160

Introduction

Affected products

A. ANACONDA LONGER LEG ILIAC STENT GRAFT SYSTEM (STRAIGHT)
B. ANACONDA LONGER LEG ILIAC STENT GRAFT SYSTEM (FLARED)
C. ANACONDA LONGER LEG ILIAC STENT GRAFT SYSTEM (TAPERED)

Reason

Vascutek has received two complaints whereby a small section of the delivery system sheath became detached from the collar during deployment of the Anaconda Longer Leg Iliac Stent device from the delivery system. Should this occur users should refer to the bailout procedure in the instructions for use (IFU) to complete the deployment.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime