UNICEL DXH 800 CELLULAR ANALYSIS SYSTEM (2017-11-06)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
SERIAL # AS23151
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
UNITED STATES
UNICEL DXH 800 CELLULAR ANALYSIS SYSTEM629029

Introduction

Affected products

UNICEL DXH 800 CELLULAR ANALYSIS SYSTEM

Reason

The shield for the single-tube station on the DxH instrument may interfere with sample aspiration in the open vial position. The mechanical interference between the shield and the right (green) position of the single-tube station can cause the tube holder to remain in the down position during aspiration and then spring up, resulting in the dislodging of an open vial. This mechanical interference may result in a biohazardous spill. The left (lavender) position tube holder of the manual station is not impacted by this issue.