DXH 500 OPEN-VIAL HEMATOLOGY SYSTEM (2017-11-06)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
1.0.2 and 1.1
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
UNITED STATES
DXH 500 OPEN-VIAL HEMATOLOGY SYSTEMB40601

Introduction

Affected products

DXH 500 OPEN-VIAL HEMATOLOGY SYSTEM

Reason

Beckman Coulter has determined that if the bleach cycle is interrupted or aborted, the DxH 500 does not provide any warning message to the operator and the system does not sufficiently flush out residual bleach prior to sample analysis. Should a bleach cycle get interrupted, the subsequent software recovery to perform a diluter reset may not be sufficient to prevent the first immediate sample run from being affected. Patient results may be impacted due to residual bleach in the diluter when a sample is processed immediately after a bleach cycle is interrupted.