Health Products

AMSCO V-PRO 60 LOW TEMPERATURE STERILIZATION SYSTEM (2017-10-12)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
s/n: 031351510
Manufacturer
Steris Mexico
790 Avante Ave., Industrial Park
Guadalupe, Nuevo Leon
67190
MEXICO
AMSCO V-PRO 60 LOW TEMPERATURE STERILIZATION SYSTEMVP40002401

Introduction

Affected products

AMSCO V-PRO 60 LOW TEMPERATURE STERILIZATION SYSTEM

Reason

STERIS has become aware that the V-PRO 60 software verifies expiration date of the Vaprox Sterilant upon completion of the sterilization cycle rather than prior to cycle initiation. Verification of the Vaprox expiration date at the end of the sterilization cycle will allow for the completion of one cycle with expired Vaprox Sterilant.