Millennium VG Variable Geometry Dual Detector Gamma Camera and VariCam (2017-10-30)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
10158
Manufacturer
GE Medical Systems Israel,
Functional Imaging
4 Hayozma Street
Tirat Hacarmel
0030200
ISRAEL
B. VARICAMASM000085
More than 10 numbers, contact manufacturer.
Manufacturer
GE Medical Systems Israel,
Functional Imaging
4 Hayozma Street
Tirat Hacarmel
0030200
ISRAEL
A. Millennium VG Variable Geometry Dual Detector Gamma CameraH2505SA

Introduction

Affected products

 

A. Millennium VG Variable Geometry Dual Detector Gamma Camera
B. VARICAM

Reason

A potential hazardous situation may occur if the collimator locking handle is not properly placed in the lock position and the warning circuit does not detect that the collimator is unlocked because of the patient proximity to the collimator during a clinical scan. There have been no injuries reported as a result of this issue.