Health Products

Carestation 620 & 650 (2017-11-03)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Datex-Ohmeda, Inc.
3030 Ohmeda Drive, P.O. Box 7550,
Madison
53707-7550
Wisconsin
UNITED STATES
B. CARESTATION 6501012-9650-000
More than 10 numbers, contact manufacturer.
Manufacturer
Datex-Ohmeda, Inc.
3030 Ohmeda Drive, P.O. Box 7550,
Madison
53707-7550
Wisconsin
UNITED STATES
A. CARESTATION 6201012-9620-000

Introduction

Affected products

A. CARESTATION 620
B. CARESTATION 650

Reason

The following issue is being addressed by GE Healthcare file no. FMI 34086: GE Healthcare has recently become aware that an incomplete seal can exist between the CO2 absorber and the breathing circuit CO2 bypass port assembly of the Carestation 600 series systems. An incomplete seal can allow rebreathing of patient gases that have bypassed the CO2 absorbent material and could result in unintended elevated levels of inspired CO2 (FICO2), which could lead to hypercarbia. There have been no injuries reported as a result of this issue.