DIMENSION VISTA SYSTEM - LOCI 8 CALIBRATOR (2017-11-07)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
7CD076
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 GBC DRIVE, MAILSTOP 514, P.O. BOX 6101
Newark
19714-6101
Delaware
UNITED STATES
DIMENSION VISTA SYSTEM - LOCI 8 CALIBRATORKC646

Introduction

Affected products

DIMENSION VISTA SYSTEM - LOCI 8 CALIBRATOR

Reason

Siemens Healthcare Diagnostics has confirmed an increase in the number of failed calibration events or negative bias with quality control (QC) and patient samples when using Dimension Vista Loci 8 Calibrator lot 7CD076 to calibrate Estradiol (E2) Assay.