AQUIOS CL FLOW CYTOMETRY SYSTEM (2017-10-02)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: Healthcare Professionals, General Public, Hospitals
Department: Health Canada
Hazard classification: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
ALL SOFTWARE VERSIONS
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
UNITED STATES
AQUIOS CL FLOW CYTOMETRY SYSTEMB30166

Introduction

Affected products

AQUIOS CL FLOW CYTOMETRY SYSTEM

Reason

When AQUIOS CL Systems are connected to a Laboratory Information System (LIS), the system may duplicate sample requests leading to sample mis-identification. For this to happen, the following conditions must be met:
  • The AQUIOS CL Flow Cytometer is connected to an LIS and
  • The AQUIOS CL System has the default test enabled and
  • Host query is enabled in LIS setup screen on system and
  • There are multiple cassettes in the autoloader and
  • The LIS host query response is received in a brief time window before LIS response timeout occurs.


When all of these conditions are met, a sample will follow two (2) paths of test request creation (LIS and Default Test).

All software versions (2.0, 2.0.1 and 2.1) in existence are impacted by this issue and all applications used on the AQUIOS CL are impacted by this issue.