When AQUIOS CL Systems are connected to a Laboratory Information System (LIS), the system may duplicate sample requests leading to sample mis-identification. For this to happen, the following conditions must be met:
The AQUIOS CL Flow Cytometer is connected to an LIS and
The AQUIOS CL System has the default test enabled and
Host query is enabled in LIS setup screen on system and
There are multiple cassettes in the autoloader and
The LIS host query response is received in a brief time window before LIS response timeout occurs.
When all of these conditions are met, a sample will follow two (2) paths of test request creation (LIS and Default Test).
All software versions (2.0, 2.0.1 and 2.1) in existence are impacted by this issue and all applications used on the AQUIOS CL are impacted by this issue.