S5I IMAGING SYSTEM and VOLCANO CORE MOBILE SYSTEM MEDICAL EQUIPMENT (2017-11-17)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: Healthcare Professionals, General Public, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots
Manufacturer
Volcano Corporation
2870 Kilgore Road
Rancho Cordova
95670
UNITED STATES
B. VOLCANO CORE MOBILE SYSTEM MEDICAL EQUIPMENT400-0100.01
All lots
Manufacturer
Volcano Corporation
2870 Kilgore Road
Rancho Cordova
95670
UNITED STATES
A. S5I IMAGING SYSTEM807400-001

Introduction

Affected products

A. S5I IMAGING SYSTEM
B. VOLCANO CORE MOBILE SYSTEM MEDICAL EQUIPMENT

Reason

Philips Volcano is initiating a voluntary correction to address a configuration issue that affects some s5i and
CORE Mobile systems with software version v3.5 (“Impacted Systems”). On certain Impacted Systems, an
unexpected Microsoft Windows Security dialog may appear during use, and the user’s response to the dialog
may adversely affect the subsequent operation of the device.
The Microsoft Windows security settings on a small number of Impacted Systems were incorrectly configured
during the manufacturing process. This misconfiguration can lead to a Windows Security Alert dialog being
displayed when the system is switched from IVUS to the FFR/iFR mode. If the user responds to the dialog by
selecting “Allow Access,” the device’s network firewall settings will be modified, opening its network ports to
potential unexpected communication from the hospital network to which the device is connected.
Unexpected communication from the hospital network could include normal network security operations, e.g.,
port scanning. If these communications occur during an active FFR/iFR procedure, data recording could be
affected leading to:
  • Incorrect FFR/iFR measurements
  • Case delay during troubleshooting and/or port scanning
  • Abandonment of use of the system


Based on our investigation there is only a remote probability that any of these impacts may occur.
A correction has been identified and put in place on the manufacturing line to prevent further occurrences of this
issue for newly manufactured systems. Based on analysis of the frequency of occurrence on the manufacturing
line, it is expected that approximately 10% of the systems running v3.5 were are potentially misconfigured.
However as this installation configuration was not recorded during the manufacturing of the device, it is unknown
which serial numbers will exhibit the Windows Security Dialog.