Health Products

NANOSTIM LEADLESS CARDIAC PACEMAKER (2017-11-17)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Hospitals, Healthcare Professionals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
Not ApplicableNANOSTIM LEADLESS CARDIAC PACEMAKER
Manufacturer
St. Jude Medical, Cardiac Rhythm Management Division
15900 VALLEY VIEW COURT
Sylmar
91342
California
UNITED STATES
S1DLCP

Introduction

Affected products

NANOSTIM LEADLESS CARDIAC PACEMAKER

Reason

As part of Abbott's post market surveillance and clinical trial monitoring processes, we have been made aware of docking button detachments that have occurred following implant or during attempted retrieval of Nanostim Leadless Cardiac Pacemaker (LCP) devices, model number S1DLCP.  In all three cases that have occurred, there has been no impact to the electrical function (e.g., pacing, sensing, communication) and no clinical impact or symptoms resulting from the docking button detachment or exposed cables.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime