All Elekta Limited Digital Accelerators with XVI or iViewGT (2017-12-06) Dec 6, 2017 · Published by Health Canada · Official notice
Share
Alert type : Medical Device Recall
Category : Medical Device
Issue : Medical Devices
Audience : Healthcare Professionals, General Public, Hospitals
Department : Health Canada
Hazard classification : Type II
Lot or serial number Companies Brand name Model or catalog number Not Applicable Manufacturer Elekta Limited Linac House, Fleming Way Crowley, Sussex RH10 9RR UNITED KINGDOM D. ELEKTA AXESSE XRT 20119 Not Applicable Manufacturer Elekta Limited Linac House, Fleming Way Crowley, Sussex RH10 9RR UNITED KINGDOM E. VERSA HD XRT 20319 Not Applicable Manufacturer Elekta Limited Linac House, Fleming Way Crowley, Sussex RH10 9RR UNITED KINGDOM B. ELEKTA INFINITY XRT 1501 Not Applicable Manufacturer Elekta Limited Linac House, Fleming Way Crowley, Sussex RH10 9RR UNITED KINGDOM C. PRECISE TREATMENT SYSTEM XRT 1083 Not Applicable Manufacturer Elekta Limited Linac House, Fleming Way Crowley, Sussex RH10 9RR UNITED KINGDOM A. ELEKTA SYNERGY PLATFORM XRT 1402
Affected products
A. ELEKTA SYNERGY PLATFORM
B. ELEKTA INFINITY
C. PRECISE TREATMENT SYSTEM
D. ELEKTA AXESSE
E. VERSA HD
Elekta has identified that if the microswitch within the PCB assembly, solenoid limit switch, of the middle arm is incorrectly setup, there is a potential for an uncontrolled extention of iViewGT / XVI detector arms.