C-450, P-600, and C625 Ceiling Lifts (2017-11-14)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
PH603850
Manufacturer
Prism Medical USA
10888 Metro Ct.
Maryland Heights
63043
Missouri
UNITED STATES
B. P-600 CEILING LIFT303090
More than 10 numbers, contact manufacturer.
Manufacturer
Prism Medical USA
10888 Metro Ct.
Maryland Heights
63043
Missouri
UNITED STATES
C. C625 CEILING LIFT323117
More than 10 numbers, contact manufacturer.
Manufacturer
Prism Medical USA
10888 Metro Ct.
Maryland Heights
63043
Missouri
UNITED STATES
A. C-450 CEILING LIFT323100

Introduction

Affected products

A. C-450 CEILING LIFT
B. P-600 CEILING LIFT
C. C625 CEILING LIFT

Reason

Prism Medical Ltd. has found that the C-Series & P600 Lift Strap for Lifts manufactured between April 25, 2017 and June 12, 2017 may wear prematurely at max load (625 lbs). They have identified the root cause as a supplied component that was not to specification. If the strap wears during use there is a potential risk for the person being supported by the lift to drop.