Health Products

INTELEPACS (2017-11-30)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: Healthcare Professionals, Hospitals, General Public
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Version 4-1-1 and up
Manufacturer
Intelerad Medical Systems Incorporated
895 De La Gauchetiere Street West, 4th floor
Montreal
H3B 4G1
Quebec
CANADA
INTELEPACSINTELEPACS

Introduction

Affected products

INTELEPACS

Reason

Following a complaint from a client, Intelerad is taking a pro-active field action to prevent occurrence of a possible patient safety problem at client sites.   The issue is triggered by the presence of the "less than" character (<) in the report. When viewing a report, the "less than" character is not present while it was there in the original report received by IntelePACS. In other cases, when the character is present in a combination with other characters, parts of the report text following the less than character are not displayed depending on the Intelerad application used by the end-user to review the reports.