Health Products

Rotating IV Pole (2017-12-12)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: Hospitals, General Public, Healthcare Professionals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Ohmeda Medical, A Division of Datex Ohmeda Inc.
8880 GORMAN ROAD
LAUREL
20723
Maryland
UNITED STATES
Rotating IV Pole6600-0851-800

Introduction

Affected products

Rotating IV Pole

Reason

GE Healthcare has recently become aware of the potential for the Rotating IV Pole to fall due to excessive tap drill depth resulting in a thin wall section of the block bracket assembly which is the bracket that attaches the pole to the bed. A fall could result in an injury to a bystander holding the patient or to the operator. There have been no injuries as a result of this issue.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime