LIFEPAK 20e DEFIBRILLATOR/MONITOR (2017-12-07)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
More than 10 numbers, contact manufacturer.LIFEPAK 20e DEFIBRILLATOR/MONITOR
Manufacturer
Physio-Control, Inc.
11811 Willows Road North East
Redmond
98052
UNITED STATES
70507-XXXXXX

Introduction

Affected Products

LIFEPAK 20e DEFIBRILLATOR/MONITOR

Reason

Physio-control is aware that some devices have had power-related failures as customers prepared their device for initial deployment or during use within the first year of distribution. The symptoms of these failures may include unexpected power on and power off, device lock-up, or a failure to power on or off, any of which has the potential to result in a failure to deliver therapy to the patient and serious injury or death. These failures are the result of manufacturing process residue located beneath a component mounted on the Power Printed Circuit Board Assembly (PCBA). There have been no adverse events reported as a result of this issue.