Boston Scientific Corporation
300 Boston Scientific Way
Marlborough
01752
Massachusetts
UNITED STATES
GUIDER SOFTIP XFH965100430

Introduction

Affected Products

GUIDER SOFTIP XF

Reason

Certain lots of Guider 7F and 8F product may be at risk of degrading within their shelf-life period. The root cause of the issue is exposure of components to UV light while in storage between 2014 and October 2017.