Health Products

ClearSight Heart Reference Sensor (2017-12-11)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer
Manufacturer
Edwards Lifesciences LLC
1 Edwards Way
Irvine
92614
UNITED STATES
CLEARSIGHT HEART REFERENCE SENSOREVHRS

Introduction

Affected Products

CLEARSIGHT HEART REFERENCE SENSOR

Reason

The current product as well as the affiliated labeling contains an expiration date of two years. Through investigation and analysis of complaint data, Edwards Lifesciences discovered that hrs units are susceptible to experiencing an "hrs out of range" error message in the EV1000 platform that are beyond six months from the date of manufacture.