After the change to standard time from Daylight Savings Time, the analyzer's time may not be consistent with local time. The analyzer date may also not reflect the current date. The discrepancy is due to a time difference between the windows operating system and the XN-L program time. After changing the time and date in the analyzer's software and cycling power to the analyzer, the date and time may not be correct when the analyzer restarts. Patient and Quality Control (QC) results are not impacted. However, the analysis date and time displayed, printed and transmitted with the results may not reflect the actual date and time the sample was run.