ARCHITECT I System - DHEA-S Assay (2017-12-19)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
00517F000
Manufacturer
Abbott GmbH and Co. K.G.
Max Planck Ring 2
Wiesbaden
65205
GERMANY
ARCHITECT I System - DHEA-S Assay8K27-20

Introduction

Affected products

ARCHITECT I System - DHEA-S Assay

Reason

Abbott has identified that falsely elevated results may be obtained when using the ARCHITECT DHEA-S Assay with samples from infants up to 60 days old. The specific cause of the elevated results is currently under investigation. Although falsely elevated results have only been observed with samples from infants up to 60 days old, the expected values representative data for children up to 10 years old provided in the ARCHITECT DHEA-S reagent insert should not be used as these reference ranges are being reviewed as part of the investigation. All in-date lots of current reagent inventory are impacted.