Health Products

Artisan - Automated Slide Stainer, Link and Link Pro (2017-12-21)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
ADL511070
Manufacturer
Dako North America, Inc.
6392 via Real
Carpinteria
93013
UNITED STATES
B. Artisan - LinkAR210
ALPB12140
Manufacturer
Dako North America, Inc.
6392 Via Real
Carpinteria
93013
UNITED STATES
C. Artisan - Link ProAR310
Not applicable
Manufacturer
Dako North America, Inc.
6392 via Real
Carpinteria
93013
UNITED STATES
A. Artisan - Automated Slide StainerAR100

Introduction

Affected Products

A. Artisan - Automated Slide Stainer
B. Artisan - Link
C. Artisan - Link Pro

Reason

A fume test has been conducted and set-up to measure the emission of formaldehyde from the Artisan instrument. Results for the fume test showed that the emission of formaldehyde exceeds the threshold limits set by the following provinces in Canada: Manitoba, Newfoundland & Labrador, Nova Scotia, Prince Edward Island.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime