D. ADVIA Centaur CP System - Tsh3-Ultra (Tsh3-Ul) Assay
06491072
Kit Lot Ending: 192, 193, 196
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
UNITED STATES
B. ADVIA Centaur System - B-Type Natriuretic Peptide (BNP) Assay
02816138
Kit Lot Ending: 298 & 301
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
UNITED STATES
C. ADVIA Centaur System - Tsh3-Ultra (Tsh3-Ul) Assay
06491072
Kit Lot Ending: 192, 193, 196
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
UNITED STATES
A. ADVIA Centaur CP System - B-Type Natriuretic Peptide (BNP) Assay
02816138
Introduction
Affected products
ADVIA Centaur CP System - B-Type Natriuretic Peptide (BNP) Assay
ADVIA Centaur System - B-Type Natriuretic Peptide (BNP) Assay
ADVIA Centaur System - Tsh3-Ultra (Tsh3-Ul) Assay
ADVIA Centaur CP System - Tsh3-Ultra (Tsh3-Ul) Assay
Reason
The ADVIA Centaur CP, XP and XPT systems will not expire the multi-diluent 15 pack at the end of the required 7-day on board stability (obs), if any of the affected B-type natriuretic peptide (BNP) or TSH3-UL lots are present on the system(s) when performing onboard dilutions. On the ADVIA Centaur XPT only, the system becomes inoperable under certain specific conditions.