Health Products

ADVIA Centaur System - B-Type Natriupretic Peptide (BNP), TSH3-ULTRA (TSH3-UL) Assay and ADVIA Centaur CP System - B-Type Natriupretic Peptide (BNP), TSH3-ULTRA (TSH3-UL) Assay (2017-12-21)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Kit Lot Ending: 298 & 301
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
UNITED STATES
D. ADVIA Centaur CP System - Tsh3-Ultra (Tsh3-Ul) Assay06491072
Kit Lot Ending: 192, 193, 196
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
UNITED STATES
B. ADVIA Centaur System - B-Type Natriuretic Peptide (BNP) Assay02816138
Kit Lot Ending: 298 & 301
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
UNITED STATES
C. ADVIA Centaur System - Tsh3-Ultra (Tsh3-Ul) Assay06491072
Kit Lot Ending: 192, 193, 196
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
UNITED STATES
A. ADVIA Centaur CP System - B-Type Natriuretic Peptide (BNP) Assay02816138

Introduction

Affected products
  1. ADVIA Centaur CP System - B-Type Natriuretic Peptide (BNP) Assay
  2. ADVIA Centaur System - B-Type Natriuretic Peptide (BNP) Assay
  3. ADVIA Centaur System - Tsh3-Ultra (Tsh3-Ul) Assay
  4. ADVIA Centaur CP System - Tsh3-Ultra (Tsh3-Ul) Assay

Reason

The ADVIA Centaur CP, XP and XPT systems will not expire the multi-diluent 15 pack at the end of the required 7-day on board stability (obs), if any of the affected B-type natriuretic peptide (BNP) or TSH3-UL lots are present on the system(s) when performing onboard dilutions. On the ADVIA Centaur XPT only, the system becomes inoperable under certain specific conditions.