Health Products

Equinox Advantage N2O/O2 Analgesic Gas Mixing and Delivery System (2017-12-21)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
EQ2150-2016
Manufacturer
O-Two Medical Technologies Inc.
45A Armthorpe Road
Brampton
L6T 5M4
Ontario
CANADA
Equinox Advantage N2O/O2 Analgesic Gas Mixing and Delivery System01EQ2000-DISS

Introduction

Affected products

Equinox Advantage N2O/O2 Analgesic Gas Mixing and Delivery System

Reason

The manufacturer had been made aware of an incident with one of their Equinox Advantage N2O/O2 mixers where the oxygen supply of the reported mixer (01EQ2000, S/N # EQ 2016-2014) was cut off due to a disengaged piston seat in the equalizer that allowed 100% supply of Nitrous Oxide to be delivered to a patient.