Health Products

XN CHECK BF (2017-12-06)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
7314
Manufacturer
Sysmex Corporation
1-5-1 Wakinohama-kaigandori
Chuo-ku
651-0073
JAPAN
XN CHECK BF213516

Introduction

Affected products

XN CHECK BF

Reason

Sysmex has become aware of an increase in recovery and imprecision of mononuclear cell counts (MN # and MN %) with the XN check BF hematology control lot 7317, level 2.