iVIEW DAB, ultraView Universal DAB, ultraView SISH DNP, OptiView DAB IHC Detection Kit and OptiView Amplification Kit, CINtec PLUS Cytology Kit, Hematoxylin II (2017-12-27) Dec 27, 2017 · Published by Health Canada · Official notice
Share
Alert type : Medical Device Recall
Category : Medical Device
Issue : Medical Devices
Audience : General Public, Healthcare Professionals, Hospitals
Department : Health Canada
Hazard classification : Type II
Lot or serial number Companies Brand name Model or catalog number More than 10 numbers, contact manufacturer. Manufacturer Roche Diagnostics GMBH Sandhoferstrasse 116 Mannheim 68305 GERMANY F. CINtec PLUS Cytology Kit 760-099 More than 10 numbers, contact manufacturer. Manufacturer Roche Diagnostics GMBH Sandhoferstrasse 116 Mannheim 68305 GERMANY G. Hematoxylin II 605-100 More than 10 numbers, contact manufacturer. Manufacturer Roche Diagnostics GMBH Sandhoferstrasse 116 Mannheim 68305 GERMANY D. OptiView DAB IHC Detection Kit 800-098 More than 10 numbers, contact manufacturer. Manufacturer Roche Diagnostics GMBH Sandhoferstrasse 116 Mannheim 68305 GERMANY E. OptiView Amplification Kit 760-700 More than 10 numbers, contact manufacturer. Manufacturer Roche Diagnostics GMBH Sandhoferstrasse 116 Mannheim 68305 GERMANY B. ultraView Universal DAB Detection Kit 760-500 More than 10 numbers, contact manufacturer. Manufacturer Roche Diagnostics GMBH Sandhoferstrasse 116 Mannheim 68305 GERMANY C. ultraView SISH DNP Detection Kit 790-2208 More than 10 numbers, contact manufacturer. Manufacturer Roche Diagnostics GMBH Sandhoferstrasse 116 Mannheim 68305 GERMANY A. iVIEW DAB Detection Kit 760-091
Affected products
iVIEW DAB Detection Kit
ultraView Universal DAB Detection Kit
ultraView SISH DNP Detection Kit
OptiView DAB IHC Detection Kit
OptiView Amplification Kit
CINtec PLUS Cytology Kit
Hematoxylin II
Related to customer complaint investigation, a potential situation concerning dispenser issues with Hematoxylin II and Horseradish peroxidase reagents where it was reported that these reagents dispensers could be leaking or sticking. This failure could result in a complete or partial dispense failure of a reagent critical to the staining reaction, potentially leading to erroneous results.