Optima CT660 and Revolution EVO (2017-06-01)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
GE Healthcare Japan Corporation
7-127 Asahigaoka, 4 Chome, Hinoshi
Tokyo
191-8503
JAPAN
B. Revolution EVORevolution EVO
More than 10 numbers, contact manufacturer.
Manufacturer
GE Healthcare Japan Corporation
7-127 Asahigaoka, 4 Chome, Hinoshi
Tokyo
191-8503
JAPAN
A. Optima CT660Optima CT660

Introduction

Affected products
  1. Optima CT660
  2. Revolution EVO

Reason

GE Healthcare has recently become aware of a potential safety issue that can result in a scan starting before the prescribed delay has completed. More specifically, if an exam is interrupted during the prep delay period using the emergency stop button, and the scan is then resumed, the delay period may be shorter than expected. If personnel are in the scan room at the time of the premature scan initiation, they can be exposed to scatter X-Ray radiation.