Health Products

Dimension Vista - Drugs of Abuse Controls (2017-12-21)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
7CQP15Dimension Vista - Drugs of Abuse Controls
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 GBC DRIVE, MAILSTOP 514, P.O. BOX 6101
Newark
19714
Delaware
UNITED STATES
KC515

Introduction

Affected products

Dimension Vista - Drugs of Abuse Controls

Reason

Siemens Healthcare Diagnostics has determined the Phencyclidine (PCP) analyte in Drugs of Abuse Positive Control [UDAT CON(+)] Lot 7cqp15 is recovering lower than expected. Customers may occasionally get a negative result for PCP with UDAT CON(+) Lot 7cqp15. All other analytes in the UDAT CON(+) recover appropriately.