C. CYTOMICS FC 500 MPL (FLOW CYTOMETER) - INSTRUMENT
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
626554
All software versions.
COULTER EPICS SERIES FLOW CYTOMETER SYSTEM - EPICS XL ANALYZER
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
626551
Introduction
Affected products
COULTER EPICS SERIES FLOW CYTOMETER SYSTEM - EPICS XL ANALYZER
EPICS XL/MCL ANALYZER
CYTOMICS FC 500 MPL (FLOW CYTOMETER) - INSTRUMENT
CYTOMICS FC 500 FLOW CYTOMETER
Reason
As a result of customer complaints and subsequent internal investigations, Beckman Coulter has determined that an internal electronic component on the circuit "amplifier" boards in the Epics XL/XL MCL Flow Cytometry Systems may be affected by a manufacturing defect. Each Epics XL/XL MCL system contains up to seven (7) of the potentially affected amplifier boards. All instruments are potentially impacted.