Health Products

Brilliance CT Big Bore System-Main Unit (2018-01-05)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Not applicable
Manufacturer
Philips Medical Systems (Cleveland) Inc.
595 Miner Road
Cleveland
44143
Ohio
UNITED STATES
Brilliance CT Big Bore System-Main Unit728244

Introduction

Affected products

Brilliance CT Big Bore System-Main Unit

Reason

Philips became aware of issues with Brilliance Big Bore CT systems running software version 4.2.0. The most significant issues are:
  • Issue #1: EFOV (extended field of view) check message does not display when FOV is set > 600mm for a paused scan
  • Issue #2: Prefer “center x/y 0” and “disable result rotation” checkboxes checked automatically when checking the ""CT Simulation""
  • Issue #3: Tumor LOC allows POI (Point of Interest) locking when a non-isocenter slice is displayed
  • Issue #4: Tracker scan completed, no images were generated
  • Issue #5: Continuous Link plans the clinical scan outside the Surview
  • Issue #6: Applying tilt on first mpr result, when the other axial results are unplanned causes change in scan length and number of cycles of scan
  • Issue #7: “System is still initializing. Please wait…” error
  • Issue #8: Auto locations not functioning as expected
  • Issue #9: System ended exam before completion
  • Issue #10: System Freeze-Up after Pause is pressed