B. Trabecular Metal Primary Hip Prosthesis Femoral Stem 128mm Stem Length Extended Offset
Manufacturer
Zimmer Inc.
1800 West Center Street
Warsaw
46580
Indiana
UNITED STATES
00-7864-013-20
63703649
A. Trabecular Metal Primary Hip Prosthesis Femoral Stem 128mm Stem Length Standard Offset
Manufacturer
Zimmer Inc.
1800 West Center Street
Warsaw
46580
Indiana
UNITED STATES
00-7864-013-00
Introduction
Affected products
Trabecular Metal Primary Hip Prosthesis Femoral Stem 128mm Stem Length Standard Offset
Trabecular Metal Primary Hip Prosthesis Femoral Stem 128mm Stem Length Extended Offset
Reason
Zimmer Biomet is conducting a medical device recall for two lots of Trabecular Metal Stems (listed). A complaint was received indicating that a "13mm extended" stem was in a box labelled as a "13mm standard" stem. The investigation determined that the lots were commingled, hence both the lots are being recalled from the field.