i-STAT Alinity Instrument with Analyzer Software (2017-12-21)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
All lots.i-STAT Alinity Instrument with Analyzer Software
Manufacturer
Abbott Point of Care Inc.
100 and 200 Abbott Park Rd.
Abbott Park
60064
Illinois
UNITED STATES
02R60-01

Introduction

Affected products

i-STAT Alinity Instrument with Analyzer Software

Reason

This product action addresses two separate issues:
  1. Abbott Point of Care (APOC) has determined that an increased frequency of quality check failure code 90-02 when testing i-STAT cartridges with the i-STAT Alinity software version Osi05 (JAMS 79A / CLEW D34).
  2. Abbott Point of Care (APOC) has also determined that when testing with i-STAT cartridges that report pH, pCO2 and pO2, an i-STAT Alinity instrument with software version OSi05 (JAMS79A / CLEW D34) that is customized to correct for user entered patient temperature may report incorrect pH and/or pCO2 results.