This product action addresses two separate issues:
Abbott Point of Care (APOC) has determined that an increased frequency of quality check failure code 90-02 when testing i-STAT cartridges with the i-STAT Alinity software version Osi05 (JAMS 79A / CLEW D34).
Abbott Point of Care (APOC) has also determined that when testing with i-STAT cartridges that report pH, pCO2 and pO2, an i-STAT Alinity instrument with software version OSi05 (JAMS79A / CLEW D34) that is customized to correct for user entered patient temperature may report incorrect pH and/or pCO2 results.