ALLURA XPER FD10/10 - Main Unit, FD20/10 System-Main Unit, FD20/20, FD10, FD20 System - Main Unit, FD20 Or Table (2017-12-28)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Not applicable
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684 PC
NETHERLANDS
F. ALLURA XPER FD20 OR TABLE000454
Not applicable
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684 PC
NETHERLANDS
D. ALLURA XPER FD10ALLURA XPER FD10
Not applicable
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684 PC
NETHERLANDS
E. ALLURA XPER FD20 SYSTEM - MAIN UNITALLURA XPER FD20
Not applicable
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684 PC
NETHERLANDS
B. ALLURA XPER FD20/10 SYSTEM-MAIN UNITALLURA XPER FD20/10
Not applicable
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684 PC
NETHERLANDS
C. ALLURA XPER FD20/20ALLURA XPER FD20/20
Not applicable
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684 PC
NETHERLANDS
A. ALLURA XPER FD10/10 - MAIN UNITALLURA XPER FD10/10

Introduction

Affected products
  1. ALLURA XPER FD10/10 - MAIN UNIT
  2. ALLURA XPER FD20/10 SYSTEM-MAIN UNIT
  3. ALLURA XPER FD20/20
  4. ALLURA XPER FD10
  5. ALLURA XPER FD20 SYSTEM - MAIN UNIT
  6. ALLURA XPER FD20 OR TABLE

Reason

Philips has become aware that extra monitors, so called 7th or 8th monitor, mounted on the Monitor Ceiling Suspension (MCS), might disengage from the MCS and fall down. The two small set screws securing the two gliding rods into an aluminum block of the support arm can come loose because of vibrations and forces exerted on them.

If the set screws have come loose gravity can keep the rods in position. However, if a force is exerted on the monitor pushing it forward, the monitor can topple forward. Once the monitor has toppled over it is only attached with its wires. If the monitor breaks loose of its wires it will drop on the floor.