Phadia 250 Instrument (including software) (2017-11-29)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
N01127
Manufacturer
PHADIA AB
RAPSGATAN 7, P.O. BOX 6460
UPPSALA
75137
SWEDEN
Phadia 250 Instrument (including software)12-3900-01

Introduction

Affected products

Phadia 250 Instrument (including software)

Reason

Erroneous results may occur as a result of a software issue when specific conditions occur with the Phadia 250. The issue can occur with all versions of the Phadia Prime Software up to and including version 2.1.4. The issue will cause software to disregard default dilution factors for rejected samples, when using the "ok to all" function for retesting of samples if specific criteria are met.