Health Products

ADIVA Centaur CP System - Folate (FOL), Folate, Cyclosporine (CSA), HBC IGM (AHBCM), DHEA-SO4 (DHEA-S) Assay and ADIVA Centaur System - HBC IGM (AHBCM), DHEA-SO4 (DHEA-S) Assay (2018-01-05)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
New York
UNITED STATES
H. ADVIA CENTAUR CP SYSTEM- CYCLOSPORINE (CSA) ASSAY04564446
All
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
New York
UNITED STATES
F. ADVIA CENTAUR SYSTEM- DHEA-SO4 (DHEAS) ASSAY06489701
All
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
New York
UNITED STATES
G. ADVIA CENTAUR CP SYSTEM- DHEA-SO4 (DHEAS) ASSAY06489701
All
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
New York
UNITED STATES
D. ADVIA CENTAUR CP SYSTEM- HBC IGM (AHBCM) ASSAY00504619
All
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
New York
UNITED STATES
E. ADVIA CENTAUR SYSTEM- HBC IGM (AHBCM) ASSAY00504619
All
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
New York
UNITED STATES
B. ADVIA CENTAUR CP SYSTEM- FOLATE ASSAY06367974
All
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
New York
UNITED STATES
C. ADVIA CENTAUR CP SYSTEM- CYCLOSPORINE (CSA) ASSAY04564446
All
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
New York
UNITED STATES
A. ADVIA CENTAUR CP SYSTEM- FOLATE (FOL) ASSAY06367974

Introduction

Affected products
  1. ADVIA CENTAUR CP SYSTEM- FOLATE (FOL) ASSAY
  2. ADVIA CENTAUR CP SYSTEM- FOLATE ASSAY
  3. ADVIA CENTAUR CP SYSTEM- CYCLOSPORINE (CSA) ASSAY
  4. ADVIA CENTAUR CP SYSTEM- HBC IGM (AHBCM) ASSAY
  5. ADVIA CENTAUR SYSTEM- HBC IGM (AHBCM) ASSAY
  6. ADVIA CENTAUR SYSTEM- DHEA-SO4 (DHEAS) ASSAY
  7. ADVIA CENTAUR CP SYSTEM- DHEA-SO4 (DHEAS) ASSAY
  8. ADVIA CENTAUR CP SYSTEM- CYCLOSPORINE (CSA) ASSAY

Reason

Siemens Healthcare Diagnostics has confirmed through internal investigation that the ADVIA Centaur Systems Cyclosporine, DHEA-s, Folate and AHBCM Assays are susceptible to biotin interference. This occurs when biotin present in patient samples interferes with the biotin-streptavidin assay architecture on the ADVIA Centaur platforms. Biotin interference has the potential to bias analytical results on these assays.