Health Products

IntelliVue Information CenteriX - Software (2018-01-08)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Not applicable
Manufacturer
Philips Medical Systems
3000 Minuteman Road
Andover
01810
Massachusetts
UNITED STATES
IntelliVue Information CenteriX - SoftwareM3290B

Introduction

Affected products

IntelliVue Information CenteriX - Software

Reason

Philips has become aware of an issue affects all PIIC iX Surveillance and Patient Link revisions. The issue will occur after a surveillance station is first restarted during the year 2018. After that initial restart, the station will thereafter no longer perform patient Discharge and Transfer operations. Subsequent application restarts will occur when any of the following operations are performed:
  • Discharging a patient under certain conditions
  • Transferring a patient (from bedside or central station)
  • Resolving patient conflicts (from bedside or central station) where bedside’s patient must overwrite PIIC iX’s patient (revision B.0x and C.0x only)
  • Clearing a surveillance sector
  • Assigning a bed to an empty surveillance sector


The surveillance station will continue to monitor patients after any restart.