Health Products

Progressa (Frame) (2018-01-12)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
S159AW9192 through S305AW2321Progressa (Frame)
Manufacturer
Hill-Rom Inc.
1069 State Route 46 East
Batesville
47006
Indiana
UNITED STATES
P7500

Introduction

Affected products

Progressa (Frame)

Reason

Progressa beds serial numbers S159AW9192 through S305AW2321 may experience a problem with the braking system. A component used within the braking system may break resulting in one or more brake pedals on the bed staying in brake or not being engaged which could result in unwanted bed movement and possible patient or user injury.