Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Not applicable
Manufacturer
Philips Medical Systems DMC GmbH
Rontgenstrabe 24
Hamburg
22335
GERMANY
ProGrade9897-010-01041

Introduction

Affected products

ProGrade

Reason

Philips has become aware that if the wifi connection between the Skyplate detector and HP transfer point is weak, an image may fail to transfer from the Skyplate detector to the system. The image remains in the memory of the detector, but cannot be transferred wirelessly or by use of the backup cable. To continue, the operator can reset the Skyplate detector by removing its batteries, but the acquired image is lost, and a re-take is necessary.