Health Products

Cortisol Saliva Luminescence Immunoassay (2018-01-12)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
LCL 130Cortisol Saliva Luminescence Immunoassay
Manufacturer
IBL International GmbH
Flughafenstrasse 52A
Hamburg
22335
GERMANY
RE62111

Introduction

Affected products

Cortisol Saliva Luminescence Immunoassay

Reason

IBL detected an issue that indicates degradation in performance of the kit that results in elevated cortisol values. When testing patient samples with one of the affected lots, the controls and calibrator values will meet specifications.