Sysmex America, Inc. has been made aware of a software defect on versions 6.0.1 and 6.0.2 running on the CellaVision DM-series and Sysmex DI-60 that may cause comments to not be transmitted to the Laboratory Information System (LIS). If all the following three conditions are met, the comments will not be transmitted from the CellaVision Software to the LIS:
The analyzer system computer, server and remote review software installed is on version 6.0.1 or 6.0.2
The workflow with multiple slides for blood samples is used
WBC (White Blood Cell), RBC (Red Blood Cell) and PLT (Platelets) comments are added to the last slide after it has been already signed