Health Products

System DI-60 Main Unit and Remote Review Software (2018-01-23)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Not applicable
Manufacturer
Sysmex Corporation
1-5-1 Wakinohama-kaigandori
Chuo-ku
651-0073
JAPAN
B. System DI-60 Remote Review SoftwareBC188562
60044 - 60965
Manufacturer
Sysmex Corporation
1-5-1 Wakinohama-kaigandori
Chuo-ku
651-0073
JAPAN
A. System DI-60 Main UnitCC286297

Introduction

Affected products
  1. System DI-60 Main Unit
  2. System DI-60 Remote Review Software

Reason

Sysmex America, Inc. has been made aware of a software defect on versions 6.0.1 and 6.0.2 running on the CellaVision DM-series and Sysmex DI-60 that may cause comments to not be transmitted to the Laboratory Information System (LIS). If all the following three conditions are met, the comments will not be transmitted from the CellaVision Software to the LIS:
  1. The analyzer system computer, server and remote review software installed is on version 6.0.1 or 6.0.2
  2. The workflow with multiple slides for blood samples is used
  3. WBC (White Blood Cell), RBC (Red Blood Cell) and PLT (Platelets) comments are added to the last slide after it has been already signed