Health Products

ARTIS Q and Zee Ceiling (2018-02-02)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
109329
Manufacturer
Siemens Healthcare GmbH
Henkestr. 127
Erlangen
91052
GERMANY
B. ARTIS Q10848281
148244
Manufacturer
Siemens Healthcare GmbH
Henkestr. 127
Erlangen
91052
GERMANY
A. ARTIS ZEE CEILING10094137

Introduction

Affected products
  1. ARTIS ZEE CEILING
  2. ARTIS Q

Reason

This update provides a corrective action in regards to the ceiling base rotation brake for select ARTIS systems. Special screws will be used for the brake adjustment. This corrective action is to address the issue with a component (adjustment screw supplied with a thread lock): a batch of non-compliant components was delivered (thread lock was missing).
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime